output: github_document
---
<!-- README.md is generated from README.rmd. Please edit that file -->
# chemCal - Calibration functions for analytical chemistry
<!-- badges: start -->
[![](https://www.r-pkg.org/badges/version/chemCal)](https://cran.r-project.org/package=chemCal)
[![Build Status](https://travis-ci.com/jranke/chemCal.svg?branch=master)](https://app.travis-ci.com/github/jranke/chemCal)
[![codecov](https://codecov.io/github/jranke/chemCal/branch/master/graphs/badge.svg)](https://codecov.io/github/jranke/chemCal)
<!-- badges: end -->
## Overview
chemCal is an R package providing some basic functions for conveniently working
with linear calibration curves with one explanatory variable.
## Installation
From within [R][r-project], get the official chemCal release using
```r
install.packages("chemCal")
```
## Usage
chemCal works with univariate linear models of class `lm`. Working with one of
the datasets coming with chemCal, we can produce a calibration plot using the
`calplot` function:
### Plotting a calibration
```r
library(chemCal)
m0 <- lm(y ~ x, data = massart97ex3)
calplot(m0)
```
![](man/figures/README-calplot-1.png)<!-- -->
### LOD and LOQ
If you use unweighted regression, as in the above example, we can calculate a
Limit Of Detection (LOD) from the calibration data.
```r
lod(m0)
#> $x
#> [1] 5.407085
#>
#> $y
#> [1] 13.63911
```
This is the minimum detectable value (German: Erfassungsgrenze), i.e. the
value where the probability that the signal is not detected although the
analyte is present is below a specified error tolerance beta (default is 0.05
following the IUPAC recommendation).
You can also calculate the decision limit (German: Nachweisgrenze), i.e.
the value that is significantly different from the blank signal
with an error tolerance alpha (default is 0.05, again following
IUPAC recommendations) by setting beta to 0.5.
```r
lod(m0, beta = 0.5)
#> $x
#> [1] 2.720388
#>
#> $y
#> [1] 8.314841
```
Furthermore, you can calculate the Limit Of Quantification (LOQ), being
defined as the value where the relative error of the quantification given the
calibration model reaches a prespecified value (default is 1/3).
```r
loq(m0)
#> $x
#> [1] 9.627349
#>
#> $y
#> [1] 22.00246
```
### Confidence intervals for measured values
Finally, you can get a confidence interval for the values
measured using the calibration curve, i.e. for the inverse
predictions using the function `inverse.predict`.
```r
inverse.predict(m0, 90)
#> $Prediction
#> [1] 43.93983
#>
#> $`Standard Error`
#> [1] 1.576985
#>
#> $Confidence
#> [1] 3.230307
#>
#> $`Confidence Limits`
#> [1] 40.70952 47.17014
```
If you have replicate measurements of the same sample,
you can also give a vector of numbers.
```r
inverse.predict(m0, c(91, 89, 87, 93, 90))
#> $Prediction
#> [1] 43.93983
#>
#> $`Standard Error`
#> [1] 0.796884
#>
#> $Confidence
#> [1] 1.632343
#>
#> $`Confidence Limits`
#> [1] 42.30749 45.57217
```
## Reference
You can use the R help system to view documentation, or you can
have a look at the [online documentation][pd-site].
[r-project]: https://r-project.org
[pd-site]: https://pkgdown.jrwb.de/chemCal